Benchtop spectrophotometers
UltraScan VIS 可见分光光度计
Benchtop instrument for laboratory use
Contact us: 021-62882816Description
高精度分光光度计,覆盖实验室工作的可见光范围。
Specifications
▶Measurement
Measurement Principle
Dual-beam benchtop spectrophotometer /colorimeter
Geometry
Diffuse d/8° reflectance, d/8° total transmission, d/0° regular transmission
Spectrophotometer/Colorimeter
Two 256 element diode arrays with a high resolution, concave holographic grating
Sphere Diameter
152 mm (6 in.)
Sphere Coating
Spectraflect™ for sphere, Duraflect™ for port plate and specular exclusion door
Port Size/Measured Area
Port Diameter/View Diameter in RSIN/RSEX reflectance modes Large Area View (LAV): 25 mm (1 in) illuminated/19 mm (0.75 in) measured Small Area View (SAV): 9.5 mm (0.375 in) illuminated/6 mm (0.25 in) measured; Port Diameter/View Diameter in TTRAN transmittance modes Large Area View (LAV): 25 mm (1 in) illuminated/17.4 mm (0.69 in) measured Small Area View (SAV): 25 mm (1 in) illuminated/10 mm (0.40 in) measured; Port Diameter/View Diameter in RTRAN transmittance mode where lens is field stop for all areas of view Large Area View (LAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured Small Area View (SAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured
Lens Switching for LAV/SAV
Automatic
Specular Component
Automated Included (RSIN) or Excluded (RSEX) in reflectance
Spectral Range
360 nm - 780 nm full CIE visible range
Wavelength Resolution
< 2 nm
Effective Bandwidth
10 nm equivalent triangular
Reporting Interval
10 nm
Photometric Range
0 to 150 %
Photometric Resolution
0.003 % (0.01 % reported)
Light Source
Pulsed Xenon lamp, filtered to approximate D65 daylight
Automatic UV Control
400 nm cutoff filter for UV control and UV exclusion; Optional 420 nm cutoff filter for UV exclusion
Flashes per Measurement
1 in LAV mode (4 in. SAV mode)
Measurement Time
< 5 seconds
Transmission Modes
Total (TTRAN) and Regular (RTRAN)
Transmission Compartment
Large and open on 3 sides, 10.2 cm D X 35.6 cm W x 16.5 cm H (4 in. D x 14 in. W x 6.5 in. H)
Standards Conformance
Reflectance: CIE 15:2004, ISO 7724/1, ASTM E1164, DIN 5033, Teil 7 and JIS Z 8722 Condition C; Transmittance: CIE 15:2004, ASTM E1164, DIN 5033 Teil 7 and JIS Z 8722 Condition E, G Haze conformance per ASTM D1003 Section 8. Procedure B Spectrophotometer
Standards Traceability
Instrument standard assignment in accordance with National Institute of Standards and Technology (NIST) following practices described in CIE Publication 44 and ASTM E259
▶Performance
Colorimetric Repeatability (20 readings range): < 0.03 ∆E* CIE L*a*b* on white tile in LAV and SAV modes; < 0.05 ∆E* CIE L*a*b* on blue denim tile in LAV and SAV modes
Spectral Repeatability
Max 0.20 peak-to-peak between 435 nm and 695 nm
Inter-instrument Agreement
∆E*; < 0.15 CIE L*a*b* (Avg) on BCRA II Tile Set ∆E*; < 0.25 CIE L*a*b* (Max) on BCRA II Tile Set
▶Physical/Electrical
Dimensions
Height: 27.9 cm (11 in.); Width: 42.0 cm (16.5 in.); Depth: 49.8 cm (19.6 in.); Weight: 20.4 kg (45 lbs)
Power
100 to 240 VAC, 47 to 63 Hz; 60 watts passive, 120 watts maximum
Interface
RS-232C serial, 19,200 baud, DB9 (female) terminal
Operating Environment
10° to 40°C (50° to 104° F), 10 % to 90 % RH, noncondensing
Storage Environment
-21° to 66°C (-5° to 150° F), 10 % to 90 % RH, noncondensing
Standard Accessories
Calibrated instrument white tile; Certificate of traceability; Black calibration light trap; Transmittance zero calibration plate; Green diagnostic tile; Wavelength diagnostic filter; Reflectance sample clamp; LAV and SAV apertures; RS-232C cable; USB-to-Serial adapter; Power cord; EasyMatch QC Software; EasyMatch QC Basic manual
Selection and learning
If you are still comparing models, scenarios, and workflows, these articles can help.
Selection guide
Benchtop vs Portable Spectrophotometers
Start from use case, sample type, and management goals to decide which instrument family fits best.
Read more->FAQ
What Is the Difference Between Benchtop and Portable Instruments?
Understand positioning, measurement method, and daily usage differences.
Read more->FAQ
What Should Be Checked During Daily Calibration?
Keep the instrument traceable and verifiable over the long term.
Read more->