Description

用于精密颜色分析和质量控制的高端台式分光光度计。

Specifications

Measurement
Measurement Principle
Dual-beam spectrophotometer/colorimeter
Geometry
Diffuse d/8° reflectance, d/8° total transmission, d/0° regular transmission
Spectrophotometer/Colorimeter
Two polychromators, each with a 512 element diode array and a high resolution, concave holographic grating
Sphere Diameter
152 mm (6 in.)
Sphere Coating
Spectraflect™ for sphere, Duraflect™ for port plate and specular exclusion door
Port Size/Measured Area
Port Diameter/View Diameter in RSIN/RSEX reflectance modes Large Area View (LAV): 25 mm (1 in) illuminated/19 mm (0.75 in) measured Medium Area View (MAV): 13 mm (0.5 in) illuminated/9 mm (0.35 in) measured Small Area View (SAV): 7 mm (0.25 in) illuminated/4 mm (0.16 in) measured; Port Diameter/View Diameter in TTRAN transmittance modes Large Area View (LAV): 25 mm (1 in) illuminated/17.4 mm (0.69 in) measured Medium Area View (MAV): 25 mm (1 in) illuminated/13.2 mm (0.52 in) measured Small Area View (SAV): 25 mm (1 in) illuminated/11.6 mm (0.46 in) measured; Port Diameter/View Diameter in RTRAN transmittance mode where lens is field stop for all areas of view Large Area View (LAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured Medium Area View (MAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured Small Area View (SAV): 17 mm (0.67 in) illuminated/17 mm (0.67 in) measured
Lens Switching for LAV/MAV/SAV
Automatic
Specular Component
Automated Included (RSIN) or Excluded (RSEX) in reflectance
Spectral Range
350 nm - 1050 nm full CIE visible range plus NIR
Wavelength Resolution
< 2 nm
Effective Bandwidth
5 nm equivalent triangular
Reporting Interval
5 nm
Photometric Range
0-150 %
Light Source
Pulsed Xenon lamps (3), calibrated and controlled in the UV range
Automatic UV Control
400 nm cutoff filter for UV control and UV exclusion; Optional 420 nm cutoff filter for UV exclusion
Transmission Modes
Total (TTRAN) and Regular (RTRAN)
Transmission Compartment
Large and open on 3 sides; 10.2 cm D X 35.6 cm W x 16.5 cm H (4 in. D x 14 in. W x 6.5 in. H)
Standards Conformance
Reflectance: CIE 15:2004, ISO 7724/1, ASTM E1164, DIN 5033, Teil 7 and JIS Z 8722 Condition C; Transmittance: CIE 15:2004, ASTM E1164, DIN 5033 Teil 7 and JIS Z 8722 Condition E, G Haze conformance per ASTM D1003 Section 8. Procedure B Spectrophotometer
Standards Traceability
Instrument standard assignment in accordance with National Institute of Standards and Technology (NIST) following practices described in CIE Publication 44 and ASTM E259
Performance
Colorimetric Repeatability (20 readings)
< 0.03 ∆E* CIE L*a*b * on white tile in LAV mode; < 0.07 ∆E* CIE L*a*b* on blue denim tile in LAV mode
Spectral Repeatability
Max 0.20 range between 435 nm and 695 nm
Inter-instrument Agreement
∆E*< 0.09 CIE L*a*b* (Avg) on BCRA II Tile Set; ∆E*< 0.20 CIE L*a*b* (Max) on BCRA II Tile Set
Physical/Electrical
Dimensions
Height: 32.3 cm (12.7 in.); Width: 42.0 cm (16.5 in.); Depth: 49.8 cm (19.6 in.); Weight: 25.9 kg (57 lbs)
Power
90 to 250 VAC, 50 to 60 Hz; 60 watts passive, 120 watts maximum
Interface
RS-232C serial, 19,200 baud, DB9 (female) terminal
Operating Environment
4° to 38°C (40° to 100° F), 10 % to 85 % RH, noncondensing
Storage Environment
-21° to 66°C (-5° to 150° F), 10 % to 90 % RH, noncondensing (-5° to 150° F)
Standard Accessories
Calibrated instrument white tile; Certificate of traceability; Black calibration light trap; Transmittance zero calibration plate; Green diagnostic tile; Wavelength diagnostic filter; Fluorescent standard; Reflectance sample clamp; LAV aperture; MAV aperture; SAV aperture; RS-232C cable; USB-to-Serial adapter; Power cord; EasyMatch QC Software; EasyMatch QC Basic manual